7 Appendix
7.1 Settings menu
Page 44
Art. no.: 2.510474 rev.: d
ARGUS LCM/PLUS
7.1.2 Special menu “About+”
V Only authorised personnel trained in the operation of this device are allowed to
do the setups in this menu.
The extended menu About+ is displayed by pressing the following button combina-
tion:
Press ; select About; press and then
Main menu extended Parameter Value
About+
Alarm Alarm stop mode On/Off
Alarm suppr. time 2 min (2 ... 10)
Alarm sound DIN EN 475/Standard
Nurse call Immediately/in 5 sec.
Alarm restriction (see page 22 ) On/Off
Param Lead II
Lead speed 12.5/25/50 mm/s
Respiration source
a
ECG/CO
2
a. When the set respiration source is ECG and the pacemaker is on, respiration measurement via the ECG is not possible. The respi-
ration wave field is not displayed in this case.
Respiration speed 6.25/12.5 mm/s
SpO
2
Pulse
4/8 or 16 s
SpO
2
sensitivity
Low /High
NIBP adult Initial 180
NIBP pediatric Initial. 150
NIBP neonatal Initial 120
NIBP MAP On
Temperature unit °C /°F
CO2 mmHg/kPA/Vol%
Trend View time 1, 5, 15, 30, 60, 120, 240 min
Device Language
English/Deutsch/
b
Français/Svenska/American/Ital-
iano/Español/Portuges/Norge
b. The Russian language requires a separate software.
Software Save as default
With this function, changed values can be perma-
nently saved. (See section 3.4.2.)
Restore defaults
With this function, the saved default values can be re-
stored. (See section 3.4.2.)
Factory defaults
Restores the SCHILLER factory settings. The display
language is changed to German. (See section 3.4.3.).
Service
For service only
NIBP update or ECG electrode test.
Program update For service only