Portable X-ray Units
Service Manual
SM-1076R2
26
5. To exit Autocalibration, press and hold Collimator Light” push-button and
then press and release “kVp Decrease” (two beeps sound). Release
“Collimator Light”.
6. Repeat the process for Small Focal Spot.
If any error message (a.e. E60) stops the Autocalibration during
the process, press “Reset” push -button and continue. If the error
persists, exit Autocalibration and repeat the Autocalibration
process from the beginning. If the problem persists, contact the
Manufacturer.
If the troubles arise at the end of the Autocalibration process (only
a few stations to end), press “Reset” push-button and continue. If
it does not work, press “Reset” push-button to store the data
corresponding to the previously autocalibrated mA stations and
start the Manual Calibration process for the r emaining stations
(Refer to section 3.2.5). If the problem still persists, contact the
Manufacturer.
IF THE D IP-SWITCH SW1-1 OF THE PORTABLE CONTROL
BOARD HAS BEEN SET IN “ON” POSITION TO ENABLE
SERVICE MODE, DO NOT FORGET TO SET IT IN “OFF”
POSITION (SERVICE MODE DISABLED) ONCE THE
CALIBRATION IS FINISHED.
Note .