Portable X-ray Units
Service Manual
SM-1076R2
43
5.2 ERROR CODES
Table 5-2
Error Code List
ERROR DESCRIPTION
No Error
Code
System does not Start-up (No indication of Activity) -- Precharge Resistors.
E01 Backup Timer I2C error.
E02 Wrong Workstation configuration.
E03 No W orkstation configured.
E04 Fluoro Order error.
E05 “Exposure” order is active during power-up.
E06 “Preparation” order is active during power-up.
E07 Wrong identifier for X-ray Tube 2.
E08 Wrong identifier for X-ray Tube 1.
E10 EEPROM corrupted or not initialized. Wrong data calibration.
E11 Load Capacitor error.
E12 mA is out of range during exposure.
E13 kVp is out of range during exposure.
E14 kVpramperror.
E15 Large Filament current is out of range.
E16 Small Filament current is out of range.
E17 DC Bus is out of range.
E19 mA without exposure order.
E20 kVp without exposure order.
E23 EEPROM error.
E24 Bucky / Digital Panel error.
E25 Large Filament current demand is over the limit.
E26 Small Filament current demand is over the limit.
E27 Collimator Light current is out of range.
E28 Collimator Light current is present without order.
E29 Generator Overload error.
E30 Wrong date/time at the RTC (Real T ime Clock).