n Setting the active preheater《ACTIVE PREHEATER》
Sets the active preheater (optional).
Input the value with the numeric keypad
and press [Enter].
Set Value Function
0 Stops controlling the active preheater.
1 Controls the active preheater (Initial value).
Setting the value to "1" causes the temperature of the active preheater to be controlled
to the set temperature of the oven. Note that turning OFF the temperature control of the
oven stops the temperature control of the active preheater.
When the oven temperature is above 85 °C, the active preheater is not controlled
at above 85 °C. Even so, you can use as passive preheater by the heat of the
oven at above 85 °C.
When the active preheater is not
installed, the panel displays [Not
Installed].
n Displaying and setting the volume of a mixer《MIXER VOL》
When a mixer information device is
connected, the panel displays the volume
of the mixer.
When a variable-volume mixer is
connected, the panel displays the screen to
select a volume of the mixer.
Input the value with the numeric keypad
and press [Enter].
A mixer volume is used for calculating the gradient delay time of the analysis system.
When no mixer information device is
connected, the panel displays [Not
Identify].
28
CTO-40C CL
3 Operation