5.6 Self Diagnostics
5-11
IRAffinity-1
The IRAffinity-1 has a self-diagnosis function. The IRAffinity-1 automatically checks its own status during
operation or startup. When all items are judged OK, the IRAffinity-1 can be used normally. The contents of the
self-diagnosis are described below.
Items checked when the IRAffinity-1 starts up or when [Measurement]-[Diagnostics] is executed:
1. Laser lighting
2. Light source lighting
3. Humidity level inside of the interferometer
4. Fixed mirror piezo actuator voltage
5. Infrared power spectrum shape
6. Presence of Auto-sampler
Items always checked while the IRAffinity-1 is operating in Status window:
1. Laser lighting
2. Light source lighting
3. Humidity level inside the interferometer
4. Status of the beam switching kit
5. Presence of Auto-sampler
6. Presence of QuickStart accessory and its information
5.6.1 Self Diagnostics during Initialization
Selecting the [Measurement]-[Initialize] initializes IRAffinity-1 and executes self diagnostics. The
result of the self diagnostics is recorded on the operation log.
When the test item passes, OK is recorded on the log. If a item failed, Fail is recorded.
Fig. 5.8 Log Screen during Initialization
5.6 Self Diagnostics
[Date & Time] Initializing: IRAffinity-1
[Date & Time] Init delay: 10 s
[Date & Time] Analyzing Piezo Voltage [101]
Setting aperture auto
[Date & Time] Piezo Voltage [101] OK
[Date & Time] Analyzing Piezo Voltage [102]
[Date & Time] Piezo Voltage [102] OK
Performing diagnostic power measurement
Setting aperture auto
[Date & Time] Power diagnostics OK (XX, XX, XX)
[Date & Time] INIT success