16 DKFD.PS.027.W3.02
2. Technical data
Technical data
SITRANS F M MAGFLO
MAG 5000 or MAG 6000 used with MAG 3100 W or MAG 1100 PFA2.6
Meter uncertainty
V: Actual flow velocity [m/s]
E: Meter uncertainty as a percentage of actual flow
MAG 6000 used with MAG 3100, MAG 1100 Ceramic or MAG 5100 W
V: Actual flow velocity [m/s]
E: Meter uncertainty as a percentage of actual flow
Reference conditions (ISO 9104 and DIN/EN 29104)
Temperature of medium 20°C ±5 K
Ambient temperature 20°C ±5 K
Supply voltage Un ±1%
Warming-up time 30 minutes
Incorporation in pipe section Inlet section 10×DN (DN ≤ 1200), 5×DN (DN > 1200)
Outlet section 5×DN (DN ≤ 1200), 3×DN (DN > 1200)
Flow conditions Fully developed flow profile
Additions in the event of deviations from reference conditions
Current output As pulse output ±(0.1% of actual flow +0.05% FSO)
Effect of ambient temperature Display/frequency/pulse output: < ±0.003% / K act.
Current output: < ±0.005% / K act.
Effect of supply voltage < 0.005% of measuring value on 1% change
Repeatability ±0.1% of actual flow for V ≥ 0.5 m/s