SITRANS F M MAGFLO
DKFD.PS.027.W3.02
6. Installation of signal converter
Installation of signal conv.
41
Locate the add-on module in the bottom of the
MAG 6000 signal converter.
6.2.1
Add-on modules
MAG 6000 only
Press the add-on module forwards as far as
possible.
The add-on module has now been installed
and the signal converter is ready to be installed
on the terminal box.
Communication to the operator menu and
electrically inputs and outputs is automatically
established by power on.
6.2.2
Remote installation -
At the sensor
Fit and connect the electrode and coil cables
as shown in chapter 7 “Electrical connections”.
The unscreened cable ends must be kept as
short as possible.
The electrode cable and the coil cable must be
kept separate to prevent interference.
Tighten the cable glands well to obtain opti-
mum sealing.
Remove the SENSORPROM
®
unit from the
sensor and mount it on the connection plate in
the signal converter.