SITRANS F M MAGFLO
3. Project guidance
DKFD.PS.027.W3.02
Project guidance
A signal converter can be supplied in a version
tested and approved for custody transfer (CT ).
The internal counter can accordingly be used
for charging.
This requires verification, sealing and setting of
the signal converter together with the sensor for
a specific flow range. After sealing the data on
the signal converter must not be changed.
The sealing of the signal converter is done by
placing sealing marks on the signal converter
and on the connection plate in the terminal box.
3.5
Custody transfer ap-
proval
3.6
Signal converter
MAG 5000 CT,
MAG 6000 CT
Sealing
The final sealing should be carried out as shown:
MAG 6000 CT is installed like a standard MAG 6000 except for the final sealing.
Calibration sealing has been carried out at calibration.