74 DKFD.PS.027.W3.02
SITRANS F M MAGFLO
9. Service
Commissioning
9. Service
Power on converter,
display light on
YES
NO
NO
Check cables/connections
Check connection board
Check pins in converter
multiplug - OK
Correct fault
YES
NO
Output
readings OK
Converter
defective
YES
Display defect
Change display
Output and display
readings OK ?
NO
NO
Check cables/connections
Check connection board
Check pins in converter
multiplug - OK
Correct fault
Error triangles flashing
NO
YES
Check error table
Converter OK -
Check settings/application
Check installation/sensor/
earthing connection etc.
YES
YES
Often problems with unstable/wrong measurements occur due to insufficient/wrong earthing or
potential equalization. Please check this connection. If OK, the SITRANS F M MAGFLO
con-
verter can be checked as described under 9.1 and sensor under 9.3.
9.1
Converter check list
When checking SITRANS F M MAGFLO
installations for malfunction the easiest method to
check the signal converter is to replace it with another MAG 5000/6000 converter with a similar
power supply.
A replacement can easily be done as all settings are stored in and downloaded from the
SENSORPROM
unit - no extra settings need to be made.
If no spare converter is available - then check converter according to check table.