Power LEDs
Located at the back of the board near the RJ 45 connector
Description Color and Meaning
Power Green – 3.3V power is available. Should be on at all times
Power Bad Red – Power is faulty or SYSCON hardware reset switch is being pressed
CAC Conn Bad Red – Connection from the SIB3 to the CAC3 is faulty or incomplete. After power up, this
LED should turn off once CAC3 to SIB3 connection is completely initialized.
I
2
C Bus LEDs, Buses A and B
Located next to I
2
C Bus connectors
Description Color and Meaning
LED2/5 Norm/Comm Dim Green – I
2
C Bus is normal
Bright Green - I
2
C Bus is communicating
LED3/6 Warning Yellow – Warning on the I
2
C Bus
LED4/7 Fault Red –I
2
C Bus fault
I
2
C Bus Pullup-Active LEDs
Located next to battery holder
Description Color and Meaning
LED19, 20 The Auto-pullup feature is supplying pullup current on the I
2
C Bus.
Can Bridge LEDs
Located to the left of the far left PCI slot
Description Color and Meaning
LED16 Ready/Comm Dim Green – Can Bridge is normal
Bright Green – Can Bridge is communicating
LED17 Warning Yellow – Warning on the Can Bridge
LED15 Fault Red – Can Bridge fault
Can I/O LEDs
Located next to far left CAN direct connector, CAN direct 5
Description Color and Meaning
LED8 TX Green On when a valid CAN I/O message (other than a heartbeat reply) has been received
and queued for processing
LED9 RX Green – On when a CAN message (other than a heartbeat transmission) has been queued
for sending to the CAN hardware
General Maintenance and Troubleshooting
4.1 Analyzer Status Indicators
Maxum edition II Analyzer General Maintenance
Manual, August 2018, A5E42019842001 23

Table of Contents

Other manuals for Siemens Maxum II

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens Maxum II and is the answer not in the manual?

Siemens Maxum II Specifications

General IconGeneral
CategoryGas Chromatograph
Measurement PrincipleGas Chromatography
Communication ProtocolsModbus, Ethernet, Profibus
Operating Temperature-20°C to 55°C
Power Supply24 VDC or 100-240 VAC
Mountingrack mount
ApplicationsRefinery, chemical, petrochemical, natural gas
DetectorsThermal Conductivity Detector (TCD), Flame Ionization Detector (FID)
Sample InletsMultiple
DimensionsVaries by configuration (e.g., 19" rack mount)