3.6 Functions
MicroSAM
Operating Instructions, 06/2012, C79000-G5376-C560-07
27
Gas flow chart for analysis modules D01 - D11 (eLive injection)
(3&
(3& 7&'%
7&'%
7&'$
7&'$
(3&
&DUULHUJDV
6DPSOH
LQOHW
6DPSOH
RXWOHW
6DPSOH
ORRS
%XIIHU
YROXPH
/LYH
LQMHFWLRQ
%DFNIOXVKLQJ
6DPSOHIORZ
/LYH
&ROXPQ
VZLWFKLQJ
7&'%7&'$
D01 ... D03
D04 ... D11
7&'$7&'%
&ROXPQ
D
&ROXPQ
E
&ROXPQ
&ROXPQ
7&'$ 7&'%
&ROXPQ
D
&ROXPQ
E
09
09
09
09
09
09
Solenoid valve 1 (MV1) Membrane valve function 1: "Bring sample loop to pressure of EPC 1"
Solenoid valve 2 (MV2) Membrane valve function 2: "Purge sample loop with sample"
Solenoid valve 3 (MV3) Membrane valve function 3: "Empty sample loop completely"
Solenoid valve 4 (MV4) Membrane valve function 4: "Inject"
Solenoid valve 5 (MV5) Membrane valve function 5: "Purge sample loop"
EPC Electronic pressure control
TCD Thermal conductivity detector
Figure 3-6 Analysis module of type D with eLive injection