Description
3.6 Functions
MicroSAM
46 Operating Instructions, 06/2012, C79000-G5376-C560-07
3.6.5.7 Cut - end
The time t
10
depends on the application: By increasing the pressure on the electronic
pressure control EPC 3, the gas components still present on column 2 are not transferred by
live switching into column 3 but are purged via TCD 2B to the outlet of the analysis.
Live with pressure EPC 3 > pressure EPC 2