3.6 Functions
MicroSAM
54 Operating Instructions, 06/2012, C79000-G5376-C560-07
Fourth temperature control system at the sample inlet
Sample path in the heated copper tube
Why an additional temperature control system?
Depending on the dew point of the process sample, condensate can form in the sample
path. This can result in destruction of the analysis module.
From the entrance to the enclosure of MicroSAM, the preheated sample is thermostatically
controlled to the correct temperature by the additional temperature control system. The
sample path is heated by the additional temperature control system.