Service and maintenance
6.1 Maintenance of analysis module
MicroSAM
86 Operating Instructions, 06/2012, C79000-G5376-C560-07
7. Attach the insulating cover.
8. Carefully screw in the cover as far as possible.
9. Rotate the 2.0 mm hexagon socket-head bolt
in the cover in the clockwise
direction until it is closed.
WARNING
Tightening the locking screw
According to the explosion protection regulations, all housing covers which have to be
unscrewed to open the housing must each be secured against unintentional opening by
a locking screw. This is achieved for this device using the hexagon socket-head bolt
.
Before opening, loosen this hexagon socket-head bolt
.
Following closing, tighten this hexagon socket-head bolt
.
10. Restart the device. During the startup process, the analysis parameters are reestablished