MOBILETT Plus HP SPR8-220.061.03 Page 14 of 22 Siemens AG
Rev. 03 09.03 CS PS 24 Medical Solutions
2 - 14 Display messages
CAL 26 Generator ON The CMOS memory
cannot be read or
programmed.
Check all supply voltages on D1
(see page 110**).
Check the voltage in the backup
battery (> 2.8 V) and the battery
jumper (see page 110**).
If all power supply voltages are
present, the error is probably on
D1.
If the message "CAL 26" occurs
repeatedly or if the message can
not be acknowledged by pressing
kV+ or by switching the unit OFF
and ON, replace D1.
CAL 27 Generator ON
during startup
Error in ROM mem-
ory. Check sum is
incorrect.
Check that all supply voltages on
D1 are present (see page 110**).
The ROM containing the program
is defective. Replace the software
(IC’s I72, I66 and I74) on D1 (see
page 110**).
If the error reoccurs after replacing
I72, I66 and I74, replace D1.
Dis-
played
messages
Mode of
operation
Cause Possible reasons and advice
* See section "Monitoring the kV and mA values"
** Wiring diagram X037E
Tab. 1

Table of Contents

Other manuals for Siemens MOBILETT Plus HP

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens MOBILETT Plus HP and is the answer not in the manual?

Siemens MOBILETT Plus HP Specifications

General IconGeneral
BrandSiemens
ModelMOBILETT Plus HP
CategoryMedical Equipment
LanguageEnglish

Related product manuals