Display messages 2 - 15
Siemens AG SPR8-220.061.03 Page 15 of 22 MOBILETT Plus HP
Medical Solutions Rev. 03 09.03 CS PS 24
CAL 28 Generator ON
during startup
An error was
detected during pro-
gram execution. The
program is inter-
rupted.
Check that all supply voltages on
D1 are present (see page 110**).
If CAL 28 reoccurs or remains
displayed, the ROM containing the
program may be defective.
Replace the software (IC’s I72, I66
and I74) on D1 (see page 110**).
If the error remains after replacing
the software, replace D1.
Enter the cause and the error
message "CAL 28" in the service
protocol.
CAL 29 Generator ON The single tank has
been connected
incorrectly.
The single tank is missing or the
cable harness has been incorrectly
connected.
Check that connection K9 on D1 is
connected.
Check for possible damage in the
main cable harness between D1
and the single tank. Measure K9:
13-4 and move the arm and single
tank while the connections are
being measured (see page 140**).
If damage is detected, replace the
main cable harness (both cables)
(see section "Replacement of the
arm cable harness").
Dis-
played
messages
Mode of
operation
Cause Possible reasons and advice
* See section "Monitoring the kV and mA values"
** Wiring diagram X037E
Tab. 1

Table of Contents

Other manuals for Siemens MOBILETT Plus HP

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens MOBILETT Plus HP and is the answer not in the manual?

Siemens MOBILETT Plus HP Specifications

General IconGeneral
BrandSiemens
ModelMOBILETT Plus HP
CategoryMedical Equipment
LanguageEnglish

Related product manuals