Siemens AG SPR8-220.061.03 Page 1 of 2 MOBILETT Plus HP
Medical Solutions Rev. 03 09.03 CS PS 24
Troubleshooting chart 3
This chart shows the sequence beginning with the generator start through to exposure
release and ends with the developed X-ray film. Potential errors are mentioned in the
function box.
Additional hints that may be helpful during troubleshooting are listed under A-E in the sec-
tion "Troubleshooting instructions".
Charge batteries
MOBILETT Plus HP
Plug in mains cable
Main switch "C"
Display illuminated
Plug out
mains cable
Main switch "B"
kV and mAs
buttons OK?
Collimator
lamp can be turned
ON/OFF
Turn plate
can be activated/
deactivated
Green
Ready light
indicator is ON
Voltage supply
Program run,
see section "P04 - Display,
LED and buzzer test"
Connections
Button function
Defective lamp
Cables
Fuse
Lamp button
remote control (option)
Cables
Fuse
Turn plate buttons
Defective magnet
Charging time
Displayed message
not acknowledged
No
No or partly
No
No
No
No
Ye s
Ye s
Ye s
Ye s
Ye s
Ye s
A
B
C
D
E
Ye s
Battery indicator
right outmost LED
active?