Loading...
MOBILETT Plus HP SPR8-220.061.03 Page 2 of 2 Siemens AG
Rev. 03 09.03 CS PS 24 Medical Solutions
3 - 2 Troubleshooting chart
* S27 is the exposure release switch. First step = PREP, second step = EXP.
s27* PREP.
Rotating anode runs
s27* EXP.
Anode speed
reached
High voltage
"Exposure"
Exposure
interrupted
Correct
density
MOBILETT Plus HP
OK
Ye s
Ye s
Ye s
Ye s
No
Ye s
Exposure release switch,
cables
and connections
Err 6 (see P02 - Rotation test)
Remote control (option)
Exposure release switch,
cables and connections
Displayed message...
Remote control (option)
kV-mAs-selection
Exposure time
Heating value
Darkroom
No
USE 39
Err 6
Err 6
No
No
No
Displayed message...
No
Displayed message...
Ye s
Displayed
message
Displayed
message