EasyManua.ls Logo

Siemens MOBILETT Plus HP - Page 92

Siemens MOBILETT Plus HP
126 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
MOBILETT Plus HP SPR8-220.061.03 Page 8 of 28 Siemens AG
Rev. 03 09.03 CS PS 24 Medical Solutions
10 - 8 Mechanical systems
Remove the upper column cover by removing the transport handle and the lower column
cover by lifting it upwards.
Raise the single tank to maximum height and secure it with a ceiling support (Fig. 10).
Remove the supporting wheels.
Put the protective mat under the unit to protect the floor from damage.
Remove battery charger D201 and motor power unit D102 (see section 7 and 11).
Disconnect all cables, ground connections and K68 according to (Fig. 11).
Remove the locking rings.
Remove the bearing rings.
Remove the shaft.
Let the turn plate down and remove it.
The ceiling suspension must be able to hold a weight of 100 kg
since the entire weight of the arm will be on the ceiling support
once the spring is removed.
If not, the arm and single tank will fall.
Make sure the ceiling support is sufficient.
Do not stand under the single tank or the arm system during the
work that follows.
Fig. 11
CAUTION
NOTE
K68
Shaft

Table of Contents

Other manuals for Siemens MOBILETT Plus HP

Related product manuals