6. Connect the two, three or four wires to four-terminal block (Ch5-RTD or Ch6-RTD) . Short-
circuit terminals as required.
&K
&K
&K57'
&K57'
',3
&K
&K
&K
&K
,VRXUFH
9VHQVH
9VHQVH
,VRXUFH
,VRXUFH
9VHQVH
9VHQVH
,VRXUFH
5666/
5666/
7(50
7(50
9'&
9'&
%
$
66/
2-wire device 3-wire device 4-wire device
Short
Isource -
Vsense -
Vsense +
Isource +
Short
Isource -
Vsense -
Vsense +
Isource +
Isource -
Vsense -
Vsense +
Isource +
7. Assemble and tighten cable gland.
4.5.5 Connecting the SSL cable
M12 connector version
The SSL cable is provided with M12 style stainless steel plugs.
The cable shield is physically and electrically terminated within the body of the plug.
Note
Never pull the cable by the plug - only by the cable itself.
1. Connect DSL using the supplied 4-wire cable with M12 connectors to both the external DSL
and the transmitter.
Note
Grounding
The SSL cable shield is only mechanically connected to the grounding terminal (PE), when
the M12 plug is correctly tightened.
Connecting
4.5Connecting the external DSL
SITRANS FS130 & FS230
52 Compact Operating Instructions, 12/2023, A5E38755607-AF