For con󹪝guration of the software parameters, see Input/output con󹪝guration (Page77).
More information can be found in the Function Manual.
The following table shows:
Which cable with which terminal
Hardware and software con󹪝guration of the channels
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
Power supply L/+ N/-
Active/+ Active/ -
Passive/+ Passive/-
Active/+ Active/ -
Passive/+ Passive/-
Active/+ Active/ - Active/+ Active/ -
Passive/+ Passive/- Passive/+ Passive/-
NO NO NO NO
NC NC NC NC
Channel 1 HART
Current output HART
Channel 2 output
Current, Frequency, Pulse and Status
Channel 1 MODBUS
Channel 1 PROFIBUS
Ch2
HW configuration
SW configuration
Power supply
Ch1
Terminals
Ch4
Ch3
In
+ (B) - (A)
Out
+ (B) - (A)
In
+ (B) - (A)
Out
+ (B) - (A)
Channel 3 & 4 Relay
Status output
Channel 3 & 4 Input and output
Outputs: Current, Frequency, Pulse and Status
Inputs: Digital
Figure4-11 Terminals/con󹪝guration overview
Ex and non-Ex versions
For Ex versions active or passive current output is preselected at the time of ordering and
cannot be changed.
Non-Ex versions can be connected as either active or passive.
4.7.2 Connecting the sensor cables (internal DSL only)
Preparing for the connection
1. Loosen the four lid screws.
2. Open the lid.
3. Remove F-connector tool from wall mount enclosure.
/
Figure4-12 F-connector tool location
Connecting
4.7Connecting wall mount enclosure
SITRANS FS130 & FS230
68 Compact Operating Instructions, 12/2023, A5E38755607-AF