4.6.5 Connecting channels 2 to 4
Connect wires
1. Remove cap and ferrule from cable gland and slide onto cable.
2. Push cable through open gland and cable path.
3. Restore ferrule and tighten cap to lightly hold cable in place.
4. Fold signal cable screen back over outer sheath and ground beneath cable clamp.
In case of shielded cables, use metal cable glands for proper connection.
5. Connect wires to terminals using screwdriver.
6. Tighten cable gland.
The numbers in the graphics below refer to table Terminal layout (industrial enclosure)
(Page56).
If connected as input or output - Industrial enclosure
Active conguration Passive conguration
⑫ IO[3] (common) ⑬ IO[3]- (passive)
⑪ IO[3]+ (active) ⑫ IO[3] (common)
Functional Earth Functional Earth
4.6.6 Connecting the power supply
1. Open power supply terminal protection cover.
2. Remove cap and ferrule from cable gland and slide onto cable.
Connecting
4.6Connecting industrial enclosure
SITRANS FS130 & FS230
Compact Operating Instructions, 12/2023, A5E38755607-AF 63