Torques
Table 7-24 Cable gland to enclosure installation torques (Siemens supplied, metric)
Description Torque (Nm)
Plastic blind plugs 1.2
Plastic cable glands 4.5
Brass cable glands/blind plugs 10
Stainless steel cable glands/blind
plugs
11
Table 7-25 Wall mount enclosure installation torques
Description Torque (Nm)
Enclosure lid screws 6
Pipe mounting nuts 10
Table 7-26 Industrial enclosure installation torques
Description Torque (Nm)
Wall bracket screws 10
Transmitter to wall bracket 25
Transmitter pedestal lock screw 10
Pedestal lock screw cap 10
Note
NPT glands on wall mount enclosure
When using NPT glands, the user must take care to use the supplied NPT thread adaptors.
7.9 Operating conditions
Table 7-27 Basic conditions
Description Speci󹪝cation
Ambient temperature
(Humidity max. 90 %)
Operation:
Transmitter without display
Display
-40 to +60 °C (-40 to +140 °F)
-20 to +60 °C (-4 to +140 °F)
Ambient temperature
(Humidity max. 90 %)
Storage: -40 to +70 °C (-40 to +158 °F)
Climate class DIN 60721-3-4
Altitude Up to 2000 m (6560 ft)
Pollution wall mount enclosure Degree 2
Technical speci󹪝cations
7.9Operating conditions
SITRANS FS130 & FS230
106 Compact Operating Instructions, 12/2023, A5E38755607-AF