A5E33329501 SITRANS LUT400 (HART) – OPERATING INSTRUCTIONS Page 101
mmmmm
General Operation
Parshall Flume
Application information
Sized by throat width
Set on solid foundation
For rated flows under free flow conditions, the head is measured at
2
/
3
the length of
the converging section from the beginning of the throat section
Note: C = Converging Dimension.
Parameter Value
2.15.1. Primary Measuring Device (PMD)
Exponential Devices
2.15.3.2. Flow Exponent
1.522–1.607
a
a.
Typical Flow Exponent range for Parshall Flume; consult your flume documen-
tation.
2.15.3.3. Maximum Head
2.15.3.4. Maximum Flow at 20 mA
2.15.4.1. K Factor
b
b.
Required for exponential device absolute calculation only.
Q
C
2
/
3
C
Zero
Head
Front View
Plan View
transducer*
Side View
* The transducer must be above the maximum head by at least the blanking value (see
2.2.4. Near
Range
).
flow