Page 210 SITRANS LUT400 (HART) – OPERATING INSTRUCTIONS A5E33329501
mmmmm
Parameters
3.3.4.6. Reminder 2 before Lifetime (Demanded)
If Remaining Lifetime (3.3.4.3.) is equal to or less than this value, the device
generates a Maintenance Demanded reminder.
a) Modify values as required.
b) Set Reminder Activation (3.3.4.4.) to the desired option.
3.3.4.7. Maintenance Status
Indicates which level of maintenance reminder is active.
In SIMATIC PDM, open the menu View – Device Status, click on the
Maintenance tab, and check the Sensor Lifetime Status window.
3.3.4.8. Acknowledged Status
Indicates which level of maintenance reminder has been acknowledged.
In SIMATIC PDM, open the menu View – Device Status, click on the
Maintenance tab and check the Sensor Lifetime Status window.
3.3.4.9. Acknowledged
Acknowledges the current maintenance reminder.
To acknowledge a reminder via SIMATIC PDM:
a) Open the menu View – Device Status and click on the Maintenance
tab.
b) In the Sensor Lifetime section click on Acknowledge Warnings.
To acknowledge a reminder via the local push buttons:
a) Navigate to Maintenance and Diagnostics (3.) > Maintenance (3.3.)
> Remaining Sensor Life (3.3.4.) > Acknowledged (3.3.3.9.), and
RIGHT arrow to acknowledge the reminder.
3.3.5. Service Schedule
The device tracks service intervals based on operating hours and monitors the
predicted lifetime to the next service. You can modify the Total Service Interval,
set schedules for maintenance reminders, and acknowledge them.
Values
Range: 0.000 to 20.000 years
Default: 0.019 years (1 week)
Notes:
• Four sets of parameters allow you to monitor the Device/Sensor Lifetimes
and set up Maintenance/Service schedules, based on operating hours
instead of a calendar-based schedule. See also Remaining Device Life
(3.3.3.), Remaining Sensor Life (3.3.4.), and Calibration Schedule (3.3.6.).
• Performing a reset to Factory Defaults will reset all the Maintenance
Schedule parameters to their factory defaults.
• The device operates in years. To view Service Interval parameters in hours
or days (only via SIMATIC PDM, PACTware FDT, and AMS) see Service
Interval (3.3.5.1.).