Page 216 SITRANS LUT400 (HART) – OPERATING INSTRUCTIONS A5E33329501
mmmmm
Parameters
3.4.1.2. Level Value
Sets the level for a fixed level simulation, or the starting level for a ramped
simulation.
3.4.1.3. Ramp
Enables/disables ramped simulation.
3.4.1.4. Ramp Rate
Sets the rate at which the simulated level will change in a ramp simulation.
3.4.2. Discrete Inputs
Simulates behaviour of external contacts connected to a discrete input.
3.4.2.1. Discrete Input 1
Disables simulation of Discrete Input 1, or sets behaviour of DI during
simulation.
3.4.2.2. Discrete Input 2
Disables simulation of Discrete Input 2, or sets behaviour of DI during
simulation.
3.4.3. Pump Activations
Sets how physical relays (that are assigned to pumps) will behave in simulation
mode.
Values
Range: Low Calibration Point to High Calibration Point
Default: 0.000
Options
*
DISABLED
ENABLED
Options
SLOW 1% of span
a
per second
a.
Low Calibration Point to High Calibration Point
*
MEDIUM 2% of span
a
per second
FAST 4% of span
a
per second
Options
*
DISABLED DI is not simulated
ON DI is simulated to be ON
OFF DI is simulated to be OFF
Options
*
DISABLED DI is not simulated
ON DI is simulated to be ON
OFF DI is simulated to be OFF
Options
*
DISABLED Pump relays are not activated in simulation
ENABLED Pump relays are activated in simulation