EasyManua.ls Logo

Sirona CEREC SpeedFire - Replacing the Main Fuses

Sirona CEREC SpeedFire
56 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
65 71 165 D3639
48 D3639.201.01.02.02 10.2016
6Care and maintenance Sirona Dental Systems GmbH
6.2Replacing the main fuses Operating Instructions CEREC SpeedFire
6.2
Replacing the main fuses
The power plug must be disconnected.
1. Use a screwdriver to screw off the 2 covers of the fuses on the rear
side of the unit counter-clockwise until the fuse holder springs out in
each case.
2. Pull out the fuse holders.
3. Replace the defective fuses (G-fuse inserts).
4. Reinsert the fuse holders.
5. Lock the covers clockwise with a screwdriver.
WARNING
Electric shock
Disconnect the power plug at the unit end before replacing the fuses.
NOTICE
Fuse type
The main fuses F1 and F2 are of the T12.5A 250V type. They are
available under the REF 65 81 891.
Use only fuses of the same type!