65 71 165 D3639
42 D3639.201.01.02.02 10.2016
5Operation Sirona Dental Systems GmbH
5.3Crystallization / glazing process IPS e.max CAD Operating Instructions CEREC SpeedFire
5.3.1 Restoration preparation for crystallization/glazing:
Fragment: CAUTION Furnace chamber size
Fragment: CAUTION Materials
NOTICE
Maximum Observe restoration size
Observe the maximum Furnace chamber size when loading the furnace.
- Diameter: 38mm
- Height: 20 mm
The restoration (including with the Glazing Support Single/Multi Unit)
must not exceed this length (38mm) and height (20mm), otherwise the
chamber may be damaged.
NOTICE
Only use approved materials
Only load the furnace with approved materials, which must be assigned
uniquely to the heat treatment job in the furnace.
NOTICE
The IPS e.max CAD Crystallization Trays and IPS e.max CAD
Crystallization Pins (Ivoclar Vivadent AG) must not be used in the
furnace as they do not fit in the combustion chamber.
Follow the additional instructions on the Ivoclar Vivadent flowchart "IPS
e.max CAD Step-by-step for CEREC SpeedFire".