6.11.12 Settings in setup mode
The parameters in setup mode allow you to customize the application program Neutral
Measurements to your specific needs.
Start setup mode by pressing the F6 Setup Mode key in the settings function level, and from there
press F6 Save to start Program Settings.
F6 Next retrieves the program settings for the highlighted application program
Display format
Decimal place
numeric
, 0 to 3
Determines the number of decimal places
alpha-numeric
, 3 digits
Determines the displayed unit of measure, e.g. lit., sqm, m
off or on
Selecting On has the effect that + and – tolerances for quantity unit control
function are specified and queried separately.
Selecting Off has the effect that the tolerance is queried only once, and is
therefore the same for + and –.
Record outside of
tolerance
within or outside
This parameter specifies whether a data record is transmitted to the printer or IT
only within the tolerance or also outside of the tolerance. If a data record is also
accepted outside of the tolerance, it is then marked with a special character.
Totaling outside of
tolerance
within or outside
This parameter specifies whether totaling into totaling memory is only
permitted for values within the tolerance or also for values outside of the
tolerance.
Delete total before
new specific weight
off or on
Selecting On has the effect that the totaling memory must first be deleted before a
new specific weight can be determined. This prevents adding different weight
types.
totaling or picking for the following settings.
means that the individual items are removed from the scale after they
have been saved to totaling memory.
means that the individual items remain on the scale after they have been
saved to totaling memory, and that the next item is placed on the scale. Factory
off or on
On means that only the respectively first item must be saved with the Plus key, all
other items are automatically saved to the totaling memory after weight has been
placed on the scale and the scale settles.