MPI-536 – USER MANUAL
10.3 Additional data
Data on additional uncertainties are useful mainly when the meter is used in
non-standard conditions and for metrological laboratories for the purpose of calibration.
10.3.1 Additional uncertainties according to IEC 61557-2 (R
ISO
)
10.3.2 Additional uncertainties according to IEC 61557-3 (Z)
1.2 m lead – 0 Ω
5 m lead – 0.011 Ω
10 m lead – 0.019 Ω
20 m lead – 0.035 Ω
WS-03 and WS-04 lead – 0.15 Ω
Mains voltage 85%..110% U
n
10.3.3 Additional uncertainties according to IEC 61557-4 (R ±200 mA)
10.3.4 Additional uncertainties of earth resistance measurement (R
E
)
Additional uncertainties according to IEC 61557-5
0% for 50 V
± 2 digits for 25 V
Serial interference voltage
Mains voltage 85%...110% U
n