MPI-536 USER MANUAL
92
3.11 Low-voltage resistance measurement
3.11.1 Resistance measurement
Select item R
X
, to call up the
measurement screen.
Select Autozero to eliminate the
resistance of test leads.
Follow the on-screen prompts.
Description of function icons
Yes accept selection
No cancels the action
After selecting Yes the meter will
give the result reduced by the
resistance of test leads.
To disable compensation of the
resistance of leads, repeat steps
with open test leads.
Then the measurement result will
contain the resistance of test
leads.
The meter is ready for measure-
ment.