EasyManua.ls Logo

Sonel MPI-536 - Page 36

Sonel MPI-536
144 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
MPI-536 USER MANUAL
34
R resistance of the tested circuit
X
L
reactance of the tested circuit
U
L-PE
voltage relative to the pro-
tective conductor
f frequency
Selecting the bar hides the
menu.
Save the measurement to the meter memory using the icon. A detailed description of
memory management is contained in section 5.3.
The last measurement can be recalled using the icon.
Double-lead measurement is not available for the UNI-Schuko adapter.
When many measurements are performed in short time intervals, the meter may
emit a large amount of heat. As a result of this, the housing of the device may be-
come hot. This is normal. In addition, the meter is equipped with the protection
against excessive temperature.
After approx. 15 consecutive measurements of the fault loop wait until the instru-
ment cools down. This limitation is due to the high current measurement and multi-
functionality of the meter.
Minimum interval between successive measurements is 5 seconds. The displayed
message indicates that it is possible to perform another measurement.
Until the message is displayed, the meter prevents any measurements.

Table of Contents

Related product manuals