MPI-536 USER MANUAL
44
Select the 3P option in the
measurement menu.
Select other settings in accord-
ance with section 3.6.1.
The meter is ready for measure-
ment.
Live mode
U interference voltage currently
on the object
Limits
R
E MAX
currently set earth re-
sistance limit
Press START to start the meas-
urement.
Read out the result.
Signal lights for reaching the limit
(section 3.6.1 step )
the result is within the set limit
the result is outside the set lim-
it
assessment not possible
After selecting the bar on the
right side, a menu will slide out
with additional measurement re-
sults.