MPI-536 – USER MANUAL
Press START to start measure-
ment.
If any of the voltages exceeds al-
lowable voltage value (50 V), Ob-
ject under voltage message is
displayed and the measurement
is blocked.
During the measurement the
H.V./REC/CONT. diode is lit in
orange.
View of the screen during meas-
urement.
The display shows the symbol of
the resistance being measured
and the progress bar of this
measurement.
The progress bar indicates the
status of the measurement.
The measurement may be can-
celled at any time using the
icon.
Read out the results.
Signal lights for reaching the limit
(section 3.7.1 step )
the result is within the set
limit
the result is outside the set
limit
assessment not possible
Additional signal lights for each of
the measured lead pairs
noise – too strong interfer-
ence signal recorded
limit – measurement taken
at inverter current limit (e.g.
short circuit in the test ob-
ject)
Save the measurement to the meter memory using the icon. A detailed description of
memory management is contained in section 5.3. The last measurement can be recalled us-
ing the icon.