NNoottee -- Make sure that the patient remains still and you do not touch the product while you weigh the patient. If the
patient moves, the scale system will try for 20 seconds to get a stable weight then
displays on the screen.
4. Press llbb//kkgg to toggle and select the scale system unit of measure (lb or kg).
WWeeiigghhiinngg aa ppaattiieenntt,, PPrriimmee CCoonnnneecctt
CCAAUUTTIIOONN
• Do not use the scale system reading as a reference for medical treatment.
• Do not use the scale for patients under 50 lb (22.7 kg).
NNoottee
• To meet the scale system accuracy claim, the patient surface must be in the horizontal position (Fowler backrest and
Gatch down) and the product cannot exceed 5 degrees of Trendelenburg or reverse Trendelenburg.
• For products with foot end push handles, make sure that you raise the foot end push handles before you operate the
scale system.
• ZZeerrooiinngg ffaaiilleedd or ZZeerrooiinngg uunnssttaabbllee may display on the screen when you zero the scale. If you are unable to zero the
scale, contact Stryker Technical Support.
• WWeeiigghhiinngg ffaaiilleedd or WWeeiigghhiinngg uunnssttaabbllee may display on the screen if the scale is not operating. If you attempt to operate
the scale again and SSeett ZZeerroo appears, contact Stryker Technical Support.
For product feature availability for your model, see
Product features
(page 19).
To weigh a patient:
1. Press and hold ZZeerroo until RReelleeaassee displays on the screen. Do not place the patient on the product until 00..00 kkgg displays
on the screen.
NNoottee -- Do not touch the product while you zero the scale system.
2. Place the patient on the product.
3. Press the WWeeiigghh button (C) to obtain the patient’s weight (
Operator keypad icon/button identification
(page 33)). The
patient’s weight displays on the screen in kg.
NNoottee -- Make sure that the patient remains still and you do not touch the product while you weigh the patient. If the
patient moves, the scale system will try for 20 seconds to get a stable weight then WWeeiigghhiinngg uunnssttaabbllee displays on the
screen.
4. Press and hold llbb//kkgg to view the patient weight in lb.
NNoottee -- The scale system unit of measure is locked in kg. When you release llbb//kkgg, the scale system unit of measure
returns to kg.
LLoocckkiinngg tthhee ssccaallee uunniitt ooff mmeeaassuurree,, PPrriimmee aanndd PPrriimmee XX
For product feature availability for your model, see
Product features
(page 19).
NNoottee -- You cannot lock the scale unit of measure for PPrriimmee CCoonnnneecctt. PPrriimmee CCoonnnneecctt defaults to display weight in kg.
When you weigh a patient, you must press and hold llbb//kkgg to view the patient weight in lb. See
Weighing a patient, Prime
Connect
(page 35).
To lock the scale unit of measure:
1. Press and hold WWeeiigghh, llbb//kkgg, and ZZeerroo until ddiiaagg displays on the screen.
2. Press llbb//kkgg or ZZeerroo to toggle to UUnniitt LLoocc.
a. For scale systems with bed exit, press and hold llbb//kkgg and ZZeerroo to select UUnniitt LLoocc.
b. For scale systems without bed exit, press and hold llbb//kkgg and ZZeerroo until rrEELL displays on the screen to select UUnniitt
LLoocc.
3. Press llbb//kkgg or ZZeerroo to toggle to llbb, kkgg, or AAnnyy.
4. Press and release llbb//kkgg and ZZeerroo. OOnnllyy and the selected unit of measure displays on the screen.
1115-309-001 Rev AD.0 35 EN

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Stryker Prime Series and is the answer not in the manual?

Stryker Prime Series Specifications

General IconGeneral
ModelPrime Series
CategoryMedical Equipment
Power Requirements120V AC, 60Hz
Battery BackupYes
TypeMedical Bed
Weight Capacity500 lbs
Featuresintegrated scale
CompatibilityCompatible with Stryker

Related product manuals