EEMMCC iinnffoorrmmaattiioonn
WWAARRNNIINNGG
• Portable RF communications equipment, including peripherals such as antenna cables and external antennas, should
be no closer than 12 inches (30 cm) to any part of the PPrriimmee SSeerriieess stretcher, including cables specified by the
manufacturer.
• Avoid stacking or placing equipment adjacent with other equipment to prevent improper operation of the product. If such
use is necessary, carefully observe stacked or adjacent equipment to make sure that they operate properly.
• The use of accessories, transducers, and cables, other than those specified or provided by the manufacturer, could
result in increased electromagnetic emissions or decreased electromagnetic immunity and result in improper operation.
The PPrriimmee SSeerriieess stretcher was evaluated using the following cables:
CCaabbllee
LLeennggtthh ((mm))
AC mains input cable
4.1
Nurse call (1/4” jack cable)
Model 1105, 1115, and 1125 PPrriimmee CCoonnnneecctt stretchers
only
3.8
GGuuiiddaannccee aanndd mmaannuuffaaccttuurreerr’’ss ddeeccllaarraattiioonn -- eelleeccttrroommaaggnneettiicc eemmiissssiioonnss
The PPrriimmee SSeerriieess stretcher is intended for use in the electromagnetic environment specified below. The customer or the
user of the PPrriimmee SSeerriieess stretcher should assure that it is used in such an environment.
EEmmiissssiioonnss tteesstt
CCoommpplliiaannccee EElleeccttrroommaaggnneettiicc eennvviirroonnmmeenntt
RF Emissions
CISPR 11
Group 1
NNoottee -- The emissions characteristics of this
equipment make it suitable for use in industrial areas
and hospitals (CISPR 11 class A). If it is used in a
residential environment (for which CISPR 11 class B
is normally required) this equipment might not offer
adequate protection to radio-frequency
communication services. The user might need to take
mitigation measures, such as relocating or re-
orienting the equipment.
RF Emissions
CISPR 11
Class A
Harmonic Emissions
IEC 61000-3-2
Class A
Voltage Fluctuations
Flicker Emissions
IEC 61000-3-3
Complies
GGuuiiddaannccee aanndd mmaannuuffaaccttuurreerr’’ss ddeeccllaarraattiioonn -- eelleeccttrroommaaggnneettiicc iimmmmuunniittyy
The PPrriimmee SSeerriieess stretcher is suitable for use in a professional healthcare facility environment and not in environments
exceeding immunity test conditions that the product was evaluated to, such as near high frequency (HF) surgical
equipment and inside of the radio frequency (RF) shielded room of magnetic resonance imaging (MRI) equipment. The
customer or the user of the PPrriimmee SSeerriieess stretcher should assure that it is used in such an environment and that the
electromagnetic environment guidance listed below is followed.
IImmmmuunniittyy tteesstt
IIEECC 6600660011 tteesstt lleevveell
CCoommpplliiaannccee lleevveell
EElleeccttrroommaaggnneettiicc
eennvviirroonnmmeenntt--gguuiiddaannccee
Electrostatic Discharge
(ESD)
IEC 61000-4-2
±8 kV contact
±15 kV air
±8 kV contact
±15 kV air
Floors should be wood,
concrete, or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30%.