NNoottee -- OOnnllyy will not display on the screen if you select AAnnyy.
5. Press and hold WWeeiigghh until qquuiitt displays on the screen.
6. Press and hold ZZeerroo until rrEELL displays on the screen. 000000..00 flashes then stops flashing.
NNoottee -- You can use the scale system when 000000..00 stops flashing.
AArrmmiinngg oorr ddiissaarrmmiinngg bbeedd eexxiitt,, PPrriimmee
Bed exit monitors the patient’s position on the product using zone controls. If the patient moves outside the limits of the
selected zone, bed exit triggers an audible alarm and an amber light illuminates from the bottom of the product litter.
For product feature availability for your model, see
Product features
(page 19).
To arm bed exit:
1. Press and hold ZZeerroo until rrEELL displays on the screen to zero the scale system. Do not place the patient on the product
until 000000..00 stops flashing.
2. Place the patient on the product.
3. Press AArrmm//DDiissaarrmm to arm zone 1. Press AArrmm//DDiissaarrmm again within 3 seconds to select and arm zone 2.
NNoottee
• The ZZoonnee 11 icon (E) or ZZoonnee 22 icon (F) illuminates to show which zone you selected (
Operator keypad icon/button
identification
(page 33)).
• You do not need to zero the scale system to change the zone or arm the bed exit again for the same patient.
To disarm bed exit, press and hold AArrmm//DDiissaarrmm until rrEELL displays on the screen.
AArrmmiinngg oorr ddiissaarrmmiinngg bbeedd eexxiitt,, PPrriimmee CCoonnnneecctt
Bed exit monitors the patient’s position on the product using zone controls. If the patient moves outside the limits of the
selected zone, bed exit triggers an audible alarm, the strip light flashes red, and an amber light flashes from the bottom of
the product litter.
NNoottee
• ZZeerrooiinngg ffaaiilleedd or ZZeerrooiinngg uunnssttaabbllee may display on the screen when you zero the scale. PPaattiieenntt rreesseett ffaaiilleedd or LLiitttteerr
uunnssttaabbllee may appear on the screen when you reset the patient information. If you are unable to zero the scale or reset
the patient information, contact Stryker Technical Support.
• BBeedd eexxiitt ffaaiilleedd, SSeett ZZeerroo, or PPaattiieenntt rreesseett rreeqquuiirreedd may display on the screen when you arm bed exit. If you are
unable to arm bed exit, contact Stryker Technical Support.
For product feature availability for your model, see
Product features
(page 19).
To arm bed exit:
1. Zero the scale system for PPrriimmee CCoonnnneecctt with scale or reset the patient information for PPrriimmee CCoonnnneecctt without scale.
a. For PPrriimmee CCoonnnneecctt with scale, press and hold ZZeerroo until RReelleeaassee displays on the screen. Do not place the patient
on the product until 00..00 kkgg displays on the screen.
b. For PPrriimmee CCoonnnneecctt without scale, press and hold the PPaattiieenntt rreesseett button (J) until RReelleeaassee displays on the screen
(
Operator keypad icon/button identification
(page 33)).
2. Place the patient on the product.
3. Press the AArrmm//DDiissaarrmm button (A) to arm zone 1 (
Operator keypad icon/button identification
(page 33)). Press the AArrmm//
DDiissaarrmm button again within 3 seconds to select and arm zone 2.
NNoottee
• The strip light illuminates green when you arm bed exit. The ZZoonnee 11 icon (E) or ZZoonnee 22 icon (F) illuminates to show
which zone you selected.
• You do not need to zero the scale system or reset the patient information to change the zone or arm the bed exit
again for the same patient.
EN 36 1115-309-001 Rev AD.0