1656 & 1657 Battery Element Tester Owner’s Manual
STS Instruments Page 54 of 88
CALibration:MEASure:VOLTage:HIGH<space><n>
Command Syntax: CALibration:MEASure:VOLTage:HIGH
<space><n>
Purpose: Sets the calibration coefficients for adjusting the voltage measurement
read-back for the tested voltage range from 2001 3000V. Available
coefficients range is 1 to 999.
Description: Sets or queries the calibration coefficients for adjusting the voltage
measurement read-back for the tested voltage range from 2001
3000V.
Parameter: 1 - 999
Query Format: CALibration:MEASure:VOLTage:HIGH?
Returns: <NR1>
CALibration:SAVe
Command Syntax: CALibration:SAVe
Purpose: Saves all three voltage measurement calibration coefficients in non-
volatile memory.
Description: All saved measurement coefficients can be retrieved using CALIBRATION
Commands.