1656 & 1657 Battery Element Tester Owner’s Manual
STS Instruments Page 53 of 88
8.4 SCPI Command Details
Instrument specific commands conform to the SCPI standard where the standard provides
guidance on the type of function that is being controlled.
8.4.1 CALIBRATION Commands
Calibration commands allow writing and reading measurement calibration coefficients to and
from the instrument.
CALibration:MEASure:VOLTage:LOW
<space><n>
Command Syntax: CALibration:MEASure:VOLTage:LOW
<space><n>
Purpose: Sets the calibration coefficients for adjusting the voltage measurement
for the tested voltage range from 300 1000V. Available coefficients
range is 1 to 999.
Description: Sets or queries the calibration coefficients for adjusting the voltage
measurement read-back for the tested voltage range from 300 1000V.
Parameter: 1 - 999
Query Format: CALibration:MEASure:VOLTage:LOW?
Returns: <NR1>
CALibration:MEASure:VOLTage:MEDium
<space><n>
Command Syntax: CALibration:MEASure:VOLTage:MEDium
<space><n>
Purpose: Sets the calibration coefficients for adjusting the voltage measurement
read-back for the tested voltage range from 1001 2000V. Available
coefficients range is 1 to 999.
Description: Sets or queries the calibration coefficients for adjusting the voltage
measurement read-back for the tested voltage range from 1001
2000V.
Parameter: 1 - 999
Query Format: CALibration:MEASure:VOLTage:MEDium?
Returns: <NR1>