EasyManua.ls Logo

STS 1656 - SOURCE Commands

Default Icon
88 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1656 & 1657 Battery Element Tester Owner’s Manual
STS Instruments Page 57 of 88
8.4.4 SOURCE Commands
Source voltage commands control the test voltage being applied to the unit under test and the
fault detection level for each test.
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude]
Command Syntax: [SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude] <n>
Purpose: Sets the test voltage level.
Description: This command sets the test voltage level. Available range is from 300 to
3000.
Query Format: [SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude]?
Returns: <NR1> Test voltage setting.
[SOURce:]VOLTage[:LEVel]:TRIGgered
Command Syntax: [SOURce:]VOLTage[:LEVel]:TRIGgered <min>, <max>
Purpose: Sets the fault detection lower (min) and upper (max) voltage trip level.
Description: This command sets lower and upper trip voltage levels for detection of a
separator fault on the unit under test. Available range for <min> and
<max> value is from 0 to 4096.
Query Format: [SOURce:]VOLTage[:LEVel]:TRIGgered?
Returns: <NR1>,<NR1> Fault detection voltage, min and max trip level settings.