1656 & 1657 Battery Element Tester Owner’s Manual
STS Instruments Page 55 of 88
8.4.2 MEASURE Commands
Measurement commands allow measurement data to be retrieved.
MEASure:VOLTage:AC?
Command Syntax: MEASure:VOLTage:AC
<space><n>
Purpose: Measures the reflected peak voltage of the reflected voltage waveform
for specified number of times <n>. This is the Q meter value shown on
the Test result screen. The <n> parameter determines how many
measurements will be taken and returned. For n=1, only a single
measurement is taken. For n >1, the response will contain <n> values
separated by commas. The maximum value for n = 100.
Description: Reads the digitized peak voltage of the reflected voltage waveform. The
engineering unit is Voltage (V).
Query Format: MEASure:VOLTage:AC?
<space><n>
Returns: <NR1>, <NR1>, <NR1>, <NR1> etc.
MEASure:VOLTage:PEAK?
Command Syntax: MEASure:VOLTage:PEAK
<space><n>
Purpose: Measures the peak of the test voltage applied to the unit under test for
the specified number of times <n>. The <n> parameter determines how
many measurements should be taken. For n=1, only a single
measurement is taken. For n >1, the response will contain <n> values
separated by commas. The maximum value for n = 100.
Description: Reads the digitized peak voltage of the voltage waveform. The
engineering unit is Voltage (V).
Query Format: MEASure:VOLTage:PEAK?
<space><n>
Returns: <NR1>, <NR1>, <NR1>, <NR1> etc.