1656 & 1657 Battery Element Tester Owner’s Manual
STS Instruments Page 70 of 88
8.7.9 *OPC?
Command Syntax: *OPC?
Purpose: This query format of the *OPC command returns a 0 when the
previously issued command has completed execution.
Description: The STS 1656/1657 executes all commands immediately so the *OPC?
query always returns a “0”.
8.7.10 *RCL <n>
Command Syntax: *RCL
<space><n>
Purpose: Recalls instrument setup from non-volatile register n.
Description: The *RCL command may be used to recall setups from non-volatile
memory. There are 10 setup registers available, numbered 1 through
10. (n = 1 ~ 10). Any active setup can be saved to a register using the
*SAV <n> command.
8.7.11 *RST
Command Syntax: *RST
Purpose: Resets instrument to power on conditions.
Description: The *RST command resets the instrument to a known default state. The
following settings will take affect after execution of a *RST command.
Parameter Reset State
INITIAL VOLTAGE
300
CONTINUOUS MODE
OFF
TRIP LEVELS
0, 500
KEYBOARD LOCK
OFF
MAX VOLTAGE
3000 V
MIN VOLTAGE
300 V
MODE
MANUAL
TEST TIME INTERVAL
1.000 SEC
OUTPUT
DISABLED
OUTPUT TESTS VOLTAGE(S)
300
8.7.12 *SAV <n>
Command Syntax: *SAV
<space><n>
Purpose: Saves instrument setup to non-volatile setup register n.
Description: The *SAV command may be used to store setups in non-volatile
memory for future recall. There are 10 setup registers available,
numbered 1 through 10. ( n = 1 ~ 10).
All setups can be recalled using the *RCL <n> command.