1656 & 1657 Battery Element Tester Owner’s Manual
STS Instruments Page 61 of 88
Error Message Description
(MEAS:VOLT:AC? 1) and cell #1 and #3 happen to fail, the
message queue will contain a 100 error message for each cell
number that failed.
100, "Test Failed
n,n,n,n,n,n"
Test performed in Auto Line mode returned at least one FAIL
result
For the Auto-Line option, when performing a measurement
(MEAS:VOLT:AC? 1) and cell #1 and #3 happen to fail, the error
message will contain PASS or FAIL indications for each cell.
Example: 100, “Test Failed FAIL, PASS, FAIL, PASS, PASS, PASS
0, "No error" No error detected
-102, "Command error" Indicates invalid command or query received
-200, "Execution error" Indicates can't execute command with parameters received
-300, "Device-specific error" Indicates instrument not properly configured
-350, "Query overflow"
A specific code entered into the queue in lieu of the code that
caused the error. This code indicates that there is no room in
the queue and an error occurred but was not recorded
-400, "Query error" Indicates query aborted
SYSTem:BEEPer:MODE
Command Syntax: SYSTem:BEEPer:MODe
<space> < off | low | high >
Purpose: Sets the loudness level of the internal buzzer used to indicate a key
press on the front panel.
Description: Three levels are available: off, low or high. For loud factory
environment, the high setting is recommended.
SYSTem:MANufacturer?
Command Syntax: SYSTem:MANufacturer?
Purpose: Returns the name of the instrument manufacturer.
Description: This command returns the manufacturer of the instrument as a
character string. This is a query format command only.
SYSTem:DATE
Command Syntax: SYSTem:DATE <space><yyyy>, <mm>, <dd>
Purpose: Sets the Real Time clock’s date to the value passed. Date format is
YYYY,MM,DD. All fields are NR1 (integer)
Description: This command sets the system date. Numeric values representing Year
(2001-2099), Month (1-12), Day (1-31). Year parameter must be 4 digits.