Daily Maintenance
Performing Shutdown 1
Before commencing Shutdown 1 confirm the following:
Make sure that slide glass has been loaded. Replenish slide glass if not sufficiently loaded.
Confirm an empty magazine is loaded in the manual magazine holder. Load an empty magazine
if it is not loaded.
Make sure that there are sufficient reagents available.
If SP-50 is on the XN track, place the analyser into ‘offline’ mode.
To perform Shutdown 1:
1. From the [Menu] select [Shutdown].