Cleaning the Stain Pool
On occasion the staining pool needs to be removed and cleaned to remove dirt or stain deposit. If using
either methanol or ethanol do not soak for longer than 30 minutes as this could damage the staining
pool.
To clean the staining pool:
1. Perform [Shutdown 1] to power down the analyser before cleaning. Instruction of how to do this
can be found in the [Daily Maintenance] section of this workbook.
2. Once the analyser has been shutdown, open the stain unit cover to gain access to the staining
pools.
3. Pull the staining pool cover down and lift and remove the staining pools. Remove both pools,
front & back.
4. Clean the staining pool by soaking it in a rinse solution (Methanol, ethanol or lukewarm water).
Make sure the staining pool is completely dry before placing back into the analyser.
5. Once dry, reinstall the staining pool and restore all affected components back to their original
position.