Sample Volumes
Sample aspiration volumes and sample volumes required for smear preparation are indicated below:
Slide
Preparation
Method
Sample
Tube
Cap
Sample Tube
Position
[Cap Open]
Aspiration
Volume
Required
Volume
Sampler
Regular
✓
Sample rack
N/A
70 µl
500 µl
RBT
✓
RBT Rack
250 µl
Manual
Regular
✓
Regular Tube
Holder
OFF
70 µl
500 µl
Regular
(1 slides)
X
ON
38 µl
300 µl
Regular
(2 slides)
X
70 µl
RBT
✓
OFF
70 µl
250 µl
Micro
(1 slide)
X
Micro Tube
Holder
N/A
38 µl
110 µl
Micro
(2 slides)
X
70 µl
140 µl