Daily Maintenance Quick Guide: Maintenance
Sysmex House, Garamonde Drive, Wymbush, Milton Keynes, MK8 8DF, United Kingdom · Phone +44 333 320 3460 · Fax +44 1908 267 901
academy.training@sysmex.co.uk· www.sysmex.co.uk
21
5. Shutdown 1 takes approximately 30 minutes depending upon the settings. The sample holder
will eject forward once the CELLCLEAN AUTO aspiration finishes. Remove and discard
CELLCLEAN AUTO tube once the sample holder ejects the empty tube. When all operations
are finished, the sample holder automatically retracts into the main unit and the instrument shuts
down.
6. Turn the instrument main power OFF using the main power switch located at the top of the
analyser.
Note: If the CELLCLEAN AUTO is not removed before shutdown finishes, a notification that a sample
tube remains in the sample holder will appear at the next startup.
7. Remove the slide glass used for cleaning. The cleaned slide will be stored in the magazine in
the manual magazine holder.
Wipe Stain Baths
1. Moisten a paper towel with methanol/ethanol
2. Carefully wipe along the top of each stain bath to remove any debris present.
IMPORTANT: The wipe down of the top of the staining bath is to only be performed when the SP-50 is
fully shutdown and is switched off to avoid the risk of injury