Monthly Maintenance – Quick Guide: Maintenance
Sysmex House, Garamonde Drive, Wymbush, Milton Keynes, MK8 8DF, United Kingdom · Phone +44 333 320 3460 · Fax +44 1908 267 901
academy.training@sysmex.co.uk· www.sysmex.co.uk
31
Monthly Maintenance
Performing Shutdown 2
Perform Shutdown 2 once a month or when instrument will not be in use for 1 week or more.
To perform Shutdown 2:
1. From the [Menu] select [Shutdown].
2. The Shutdown dialog box appears and the manual tube holder extends. Select [Shutdown 2].