As Required Maintenance – Quick Guide: Maintenance
Sysmex House, Garamonde Drive, Wymbush, Milton Keynes, MK8 8DF, United Kingdom · Phone +44 333 320 3460 · Fax +44 1908 267 901
academy.training@sysmex.co.uk· www.sysmex.co.uk
6. Input the new reagent information by scanning the barcode on the outer box/container of the
new reagent. You can also enter the reagent code manually by selecting the reagent required
and entering the information using the onboard keypad. If changing manually, make sure the
[Replace the reagent] box is ticked.
7. When the correct reagent information has been entered the red warning triangle will be replaced
by a green tick.
8. Remove the old container and replace it with the new reagent. Select [Execute] to prime the new
reagent. Once complete the [Reagent Replacement] dialog box closes.