Preparing Samples
5–8 TA I
NSTRUMENTS
DSC A
UTOSAMPLER
CE
Figure 5.7
Installing the
New Inner Core
8. Insert the modified die into the sample press.
9. Tilt the sample press on its back again, and
turn the base screw clockwise to adjust the
height of the lower die.
10. Tilt the sample press back up, and bring the
lever down to test the height of the lower
die. The upper die should just touch the
lower die when the lever is fully down.
11. Repeat steps 9 and 10 until the height of the
lower die is satisfactory. Then tighten the
lower die holder thumbscrew to hold the
lower die in place.
12. Crimp a few empty pans to verify that the
die is at an acceptable height (see pages 5-
19 to 5-23).